Contact an Account ManagerPCB from sludge
Summary
Type: SPE
Application #: 301400
Catalog#: 730149 - SPE Cartridges, Chromabond, NAN, 6ml, PP, 700/2000/700mg 30/pk
Detection: For further analysis we recommend GC
e.g. with capillary column OPTIMA 5, 0.25 µm film, 50 m x 0.25 mm ID.
Conditions: Column conditioning:
10 ml n-hexane
Sample application:
Aspirate 2 ml of the extract into the column
Elution:
Aspirate 40 ml n-hexane slowly through the column with weak vacuum, then
evaporate and fill to 5 ml with n-hexane
Authors: SPE department
Source: Macherey-Nagel
Details
Legend: Compound recovery [%]
PCB 28 104
PCB 52 100
PCB 101 99
PCB 138 98
PCB 153 101
PCB 180 98
PCB 209 104